Large Scanning TCT System

Large Scanning-TCT aims for the same applications as Scanning TCT. The main difference with respect to the latter is the vertical placement of the optical system and larger and more robust stages that can carry heavy load.

The mounting surface is cooled by a combination of several Peltier elements with the combined power of 200W. The mounting plane is much larger and allows also for wider range of movements. Performance wise the properties of both systems are similar.

It can be used for standard and Edge-TCT measurements. Edge-TCT enables measurements of charge collection and carrier velocity profiles– crucial for understanding the performance of heavily irradiated detectors.

Technical Specifications

Component

Parameter

Properties

Laser

Laser diode
Wavelength 660 nm or 1064 nm (optional others on demand)
Pulse power Few M.I.P. – 100 M.I.P. (equivalent in 300 micron Si)
Pulse width <350 ps – 4000 ps (tunable)
Coupling Single core fibre

Driver

Control With PC over USB
Running mode Single pulse: 50 Hz – 1 MHz
Pattern mode mHz to 100 kHz
Sequence 1024 bits deep sequence of pulses
Minimum distance between pulses 440 ns
External control External NIM logical signal
I/O Ext. trigger in / Trigger out MC / Trigger out laser

Optics

Beam spot (FWHM) <11 μm @1064 nm, <8 μm @660 nm
Coupling Fibre coupled
Attenuation Through collimator on beam-expander

Mechanics

Experimental box 78 × 60 × 60 cm (H × W × L)
Translation 3-axis computer controlled
Load / stage 50 kg
Moving range 10 × 10 × 10 cm (X, Y, Z)
Position resolution <1 μm
Control Via USB

Electronics

Amplifier 53 dB
Bias-T range >1000 V
Frequency range <0.3 MHz – >3000 MHz

Mounting table

Cooling type Peltier element
Heat remover Water / inlets provided
Cooling power 200 W (4 × 50 W) – ΔT >40 °C
Test sample box 2.5 × 3 cm² – testing box
Mounting plane 20 × 20 cm²

Software

Full DAQ Adaptation for specific oscilloscope, power source on demand
Analysis software ROOT based package